GE Healthcare Innova 4100 Digital Fluoroscopic Imaging (Cardiovascular Imaging System

FDA Recall Z-1522-2008 · Posted 2008-09-30

Recalling FirmDatex - Ohmeda, Inc
Recall StatusTerminated
Product CodeMQB
Date Initiated2008-05-19
Date Terminated2016-06-30
Product Quantity87

Reason for Recall

GE Healthcare has become aware of a potential condition that may impact operator or patient safety. One customer has reported that inability to terminate X-ray exposure after releasing the hand switch control (located in the control room).

Firm Action

An Urgent Medical Device Correction letter, dated June 2008 was sent to customers The letter described the issue, affected products and provided short term user recommendations. GE developed a new software release that corrects this condition and will launch the corrective action to upgrade all affected systems.

Root Cause

Software design

Distribution Pattern

Worldwide Distribution - AL, AZ, CA, FL,, GA, HI, IA,IL, IN, KY, LA,MA, MD, MI, MO, MS, NC, NH, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WI, and WV, Puerto Rico. OUS to include: Australia, Byelarus, China, Denmark, France, Germany, Greece, Honduras, Iran, Iraq, Italy, Japan, Latvia, Lebanon, Mexico, Poland, Russia, Saudi Arabia, Turkey, United Kingdom, and Venezuela.