Shiley Specialized Tracheostomy Tube, M6DCT, Sterile EO, Made in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA. Intended for bypass of upper airway obstructions, provide long term ventilation, support and/or manage tracheal/bronchial secretions.

FDA Recall Z-1522-2010 · Posted 2010-05-06

Recall StatusTerminated
Product CodeJOH
Date Initiated2010-04-13
Date Terminated2012-01-18
Product Quantity2 units

Reason for Recall

Inflatable cuff Tracheostomy tubes may leak air, resulting in inadequate ventilation.

Firm Action

Covidien issued an "Urgent Medical Device Recall" notification dated April 13, 2010. Consignees were alerted to the affected device. If an affected product is already in use, the firm recommends that the tube be replaced as soon as the patient's physician determines it is clinically appropriate. All affected product is to be returned to the firm. For further Information, contact Covidien at 1-800-635-5267.

Root Cause

Process design

Distribution Pattern

Worldwide Distribution -- United States (including Guam and Puerto Rico), Australia, Belgium, Brazil, Canada, Chile, Colombia, Honduras, Japan, Mexico, Panama, Singapore and Uruguay.