Open, Classified

CRT-P Quad Models (CRT-Ps): a. Percepta CRT-P MRI: W1TR01, W1TR04, W4TR01, W4TR04 b. Serena CRT-P MRI: W1TR02, W1TR05, W4TR02, W4TR05 c. Solara CRT-P MRI: W1TR03, W1TR06, W4TR03, W4TR06. implantable medical devices used to treat arrhythmias related to bradycardia, and cardiac resynchronization therapy (CRT)

FDA Recall Z-1523-2021 · Posted None

Recall StatusOpen, Classified
Product CodeNKE
Date Initiated2021-04-06
Product Quantity45,411 devices

Reason for Recall

There is a SmartSync software issue that results in an overestimation in the displayed longevity of these devices during an approximate 6-month window of time before the device triggers its Recommended Replacement Time (RRT).

Firm Action

Beginning 06-APR-2021, Medtronic notified risk managers and practice managers of the field corrective action notification via 2-day UPS delivery of the consignee letter. Confirmation delivery of the consignee letter will be obtained. Medtronic Field Representatives have been instructed to begin installation of the software update. Outside the United States (OUS), communication began on 06- APR-2021 using locally approved methods. Medtronic Field Representatives will begin installation of the software update as it is available locally. Consignees were advised that Medtronic representatives can assist in installing the software update on programmers. During device interrogation with an updated programmer at implant or the next regularly scheduled clinic visit, a patients device will automatically receive the updated software. A courtesy notification will be provided to physicians. No confirmation will be required for the courtesy notification.

Root Cause

Software design

Distribution Pattern

Worldwide

Official Source

View FDA Recall →