Normed TX 8 Driver, Non-Cannulated (used with the Ascension Total Foot System); Catalog #503004268; distributed by Ascension Orthopedics, Inc., Austin, TX 78754. Orthopedic manual surgical screwdriver.
FDA Recall Z-1524-2009 · Posted 2009-06-26
Advertisement
Recalling FirmAscension Orthopedics, Inc
Recall StatusTerminated
Product CodeHXX
Date Initiated2009-04-14
Date Terminated2010-09-24
Product Quantity9 units.
Reason for Recall
Screwdriver used with the Ascension Total Foot System not manufactured to specification, resulting in the driver not properly engaging and/or stripping the screws making it difficult to properly seat the screws during surgery.
Firm Action
Firm notified their distributors of recall via phone starting 04/14/09. Distributors replaced the recalled screwdrivers with ones that were manufactured to specification.
Root Cause
Pending
Distribution Pattern
Nationwide Distribution -- including states of CO, OR, and TX.