BARD Biopsy Systems UltraClip Dual Trigger breast tissue marker. Intended for use to attach to soft breast tissue at the surgical site during an open surgical breast biopsy.

FDA Recall Z-1524-2013 · Posted 2013-06-08

Recall StatusTerminated
Product CodeFZP
Date Initiated2013-04-16
Date Terminated2014-01-30
Product Quantity3,085

Reason for Recall

Bard Peripheral Vascular (BPV) has confirmed that some devices with the recalled product code / lot number combination may have the metal tissue marker separated from the needle prior to use.

Firm Action

Bard Peripheral Vascular, Inc. initiated a recall of certain product code and lot number combinations of Bard UltraClip Dual Trigger Tissue Marker by contacting customers on 04/16/2013. On 04/18/2013, BPV sent a Customer Recall notification Letter via FedEx with proof of delivery notice. The letter included reason for recall, a table of affected product code and lot numbers, instructions (Do not use or further distribute any affected product), a response form (titled Recall and Effectiveness Check Form) enclosed, and contact information.

Root Cause

Nonconforming Material/Component

Distribution Pattern

Worldwide Distribution: USA Nationwide and Internationally to Europe.