Terminated

PowerPort isp M.R.I Implanted Port without Suture Plugs with attachable 6F Chronoflex Polyurethane Open-ended Single -Lumen Venous Catheter, BARD Access Systems Assembled in Mexico Indicated for patient therapies requiring repeated access to the vascular system for the infusion of medications, I.V. fluids, parenteral nutrition solutions, blood products and blood sample withdrawals.

FDA Recall Z-1525-2010 · Posted 2010-04-29

Recalling FirmBard Access Systems
Recall StatusTerminated
Product CodeLJT
Date Initiated2010-03-18
Date Terminated2010-10-01
Product Quantity200 units total (100 units of each lot, see code information)

Reason for Recall

8F PowerPort stems were mixed with 6F PowerPorts and packaged.

Firm Action

The firm, Bard Access Systems, Inc. (BAS) sent a "URGENT: Powerport isp M.R.I with 6F Chronoflex Catheter RECALL" letter dated March 18, 2010, to the customers. The letter describes the product, problem and actions to be taken by customers. The customers were instructed to immediately examine their inventory and quarantine all products subject to this recall and notify any customers they may have distributed the product to; DO NOT USE THIS PRODUCT OR FURTHER DISTRIBUTE ANY OF THE IMPLICATED LOTS; remove and return the affected lots to BAS, and complete and Fax the Reply Form and Inventory Reconciliation Form, if unable to Fax call and report information. Only unused products, in its original packaging, should be returned to BAS. Should you have any questions or require assistance in this matter, please contact BAS Customer Service by calling our toll-free number 1-800-290-1689.

Root Cause

Mixed-up of materials/components

Distribution Pattern

Nationwide distribution: AZ, CA, DC, FL, MA, MD, NC, NE, NV, NY, PA and SC.