Terminated

Clinical Users Guide (CUG) accompanying the EpiFLO SD, Part #01-100-1000000. The EpiFLO is packaged in a sealed poly-bag that contains: 1) one sterile tyvek packaged pouch containing the cannula, 2) one non-sterile EpiFLO generator, and 3) IFU (instructions for use). Five (5) of these packages are then placed in a cardboard box. There is an Ogenix label on the poly bag, on the EpiFLO generator, and on the cardboard box. The EpiFLO SD is intended to provide transdermal sustained oxygen delivery.

FDA Recall Z-1526-06 · Posted 2006-09-28

Recalling FirmOgenix Corporation
Recall StatusTerminated
Product CodeKPJ
Date Initiated2006-03-24
Date Terminated2007-12-19
Product Quantity518

Reason for Recall

Conflicting instructions are provided in the Instructions for Use (IFU) and the Clinical Users Guide (CUG) which accompany the EpiFLO device. Consequently, the firm removed the CUG from distribution.

Firm Action

Ogenix began notifying their customers of the recall via telephone on 3/24/2006. All thirteen (13) customers were initially contacted via telephone. Those customers which could not be reached were contacted by mail. Three (3) attempts were made via telephone with the fourth attempt being made by letter until all of the customers were reached. The customers were instructed to destroy all CUGs in their possession and the distributor/customers were asked to contact any of their customer(s) who may have received the incorrect CUGs and request that they destroy the brochures.

Root Cause

Other

Distribution Pattern

The product was distributed to customers located in the following states: IN, MD, TX, MA, MI, MN, OH, FL, and NB.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →