Recall StatusTerminated
Date Initiated2010-03-12
Date Terminated2012-01-30
Product Quantity18 (4 USA, 14 OUS)
Reason for Recall
Unanticipated shut-down: A specific lot of GE Healthcare Aisys and Avance machines control board wiring harness have a defect, which can cause the machine to unexpectedly shut down, terminating ventilation, anesthetic delivery, and potentially patient monitoring.
Firm Action
Consignees were sent a GE Healthcare "Urgent Medical Device Correction" letter dated March 12, 2010. The letter was addressed to Chief of Anesthesia, Health Care Administrator/Risk Manager, Director of Biomedical/Clinical Engineering.
The letter listed the Safety Issue, Affected Product Details, Safety Instructions, Product Correction and Contact Information.
Distribution Pattern
Worldwide distribution: USA (CA, FL, TN), CANADA, DENMARK, FINLAND, FRANCE, ITALY, SAUDI ARABIA, SOUTH AFRICA, SPAIN, UNITED ARAB EMIRATES, and UNITED KINGDOM.