PolypVac 3.5mm Microdebrider; Catalog Number FG-000016; contents: 1 Microdebrider, 1 Stylet, 1 filter. Sterile. Used for the excision of polyps in the nasal passageways and in sinuses for which access has been previously established.
FDA Recall Z-1527-2014 · Posted 2014-04-30
Advertisement
Recalling FirmLaurimed LLC
Recall StatusTerminated
Product CodeERL
Date Initiated2014-04-07
Date Terminated2014-05-14
Product Quantity58
Reason for Recall
The reciprocation mechanism in the PolypVac Microdebrider may turn off prematurely due to a component failure in the device handle.
Firm Action
All direct accounts were notified on April 8, 2014 with an Urgent: Medical Device Recall of PolypVac Microdebrider letter, sent by first class mail. The letter identified the affected product, the reason for the recall, and stated that the firm is in the process of implementing corrections. The firm is requesting that all unused lots be returned to them. Replacement devices will sent to customers. In addition, customers are to complete the attached Lot Information sheet. Questions should be directed to [email protected] or 650-587-5296.
Root Cause
Component design/selection
Distribution Pattern
Distributed in the states of CA, VT, NC, DO, TX, FL, PA, OR, AZ, IL, ID, and VA.