Recall StatusTerminated
Date Initiated2020-02-14
Date Terminated2024-06-12
Product Quantity762
Reason for Recall
Embolization device delivery system may fracture at the distal section during device implantation. A weakened bond may result in unintended separation, where the distal portion of the device delivery system remains in the patient. If a Pipeline Flex embolization device has already been implanted successfully, there is no increased risk to patients due to this issue.
Firm Action
On 02/14/20, Medtronic mailed Urgent Medical Device Recall notices to Healthcare Professionals and Risk Managers. They were advised of the following:
1) Do not use any affected product. Remove and quarantine all unused affected products in your inventory.
2) Return the affected products. Medtronic representatives can assist with facilitating the return of affected product and with identifying suitable replacement product.
3) Complete and return the Customer Confirmation Form.
In addition, customers were asked to share the communication within their organization, with other organizations where affected devices were transferred, and any other associated organizations that may be impacted.
On 09/04/2020, a Safety and Labeling Notification was sent to customers informing them that that Instructions for Use will be supplemented to update the Precaution and Warning statements and the Device Complications section. Medtronic recommends that physician users be attentive to the updated information in the Instructions for Use and include this information in discussions with patients. Further, please share this communication within your organization, with other organizations where these devices have been transferred, and any other associated organizations that may be impacted by this action.
Root Cause
Process change control
Distribution Pattern
Worldwide distribution. US nationwide, Argentina, Australia, Austria, Brazil, Bulgaria, Canada, China, Colombia, Croatia, Czech Republic, Denmark, Ecuador, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Italy, Japan, Jordan, Kazakhstan, Korea, Lebanon, Mexico, Netherlands, Norway, Paraguay, Poland, Qatar, Russian Federation, Saudi Arabia, Serbia, Slovakia, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, Uruguay, and Viet Nam
Official Source