Cytomics FC 500 Flow Cytometry System with Data Innovations Instrument Manager, Part Numbers 626553 and 6605628 A system for the qualitative and quantitative measurement of biological and physical properties of cells and other particles.
FDA Recall Z-1529-2008 · Posted 2008-08-08
Advertisement
Recalling FirmBeckman Coulter Inc
Recall StatusTerminated
Product CodeGKZ
Date Initiated2007-08-20
Date Terminated2012-10-15
Product Quantity141 units
Reason for Recall
Labeling provided by two integrated software systems are not clear enough to avoid potential demographic and sample type mismatches under certain conditions.
Firm Action
A Product Corrective Action (PCA) letter, dated August 16, 2007, informed the customers that the User documentation and Online Help for the FC500 with Data Innovations lnstrument Manager are not clear enough to avoid potential demographic and sample type mismatches under certain conditions. The letter asks the customers to share the information with their laboratory staff and retain the notification as part of their laboratory Quality System documentation. They are also to review all sample identification, demographics, data and results before reporting results out. A response form is included. Questions and concerns regarding the notification should be directed to Beckman Coulter Customer Service at 1-800-526-7694.
Root Cause
Software design
Distribution Pattern
Worldwide Distribution --- including USA and country of Canada.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.