Terminated

Alfa Wassermann Diagnostic Technologies, LLC. ACE ALT Reagent. Product Usage: For quantitative determination of ALT in serum and lithium heparin plasma using ACE, ACE Alera and ACE Axcel Clinical Chemistry Systems. This test is intended for use in clinical laboratories and physician office labs.

FDA Recall Z-1529-2014 · Posted 2014-04-30

Recalling FirmAlfa Wassermann, Inc.
Recall StatusTerminated
Product CodeCKA
Date Initiated2014-03-05
Date Terminated2015-08-13
Product Quantity731 Kits

Reason for Recall

Extremely low absorbance readings.

Firm Action

Alfa Wassermann Diagnostic Technologies, LLC issued an "Important: Advisory Notice/"Product Response Form" dated March 5, 2014 and sent on March 6, 2014 to their customers for ACE ALT Reagent Product reorder number SA1046 Lot F3604. They issued a follow-up "Important: Advisory Notice/"Product Response Form" dated March 21, 2014 and sent on March 26, 2014 to their customers for ACE AST Reagent Product reorder numbers SA1047, Lot F3609, and RX1047, Lot F3610. The notification identified the product(s), problem, and actions to be taken by the customers. Customers were instructed to destroy any affected product and complete and submit the enclosed Product Response Form. For questions or concerns, contact the Alfa Wassermann Diagnostic Technologies, LLC Customer Solution Center, Toll Free: (866) 419-ALFA (2532).

Root Cause

Process control

Distribution Pattern

US Nationwide Distribution