Accelerate PhenoTest BC kit, Ref #10101018. The firm name on the label is Accelerate Diagnostics, Tucson, AZ. A multiplexed in-vitro diagnostic test intended for use with the Accelerate Pheno system.

FDA Recall Z-1529-2018 · Posted None

Recall StatusTerminated
Product CodePRH
Date Initiated2018-03-14
Date Terminated2023-01-17
Product Quantity95 kits

Reason for Recall

The kits may produce an elevated rate of false positive calls for the Staphylococcus aureus identification probe (SAU).

Firm Action

The recalling firm contacted all customers on 3/14/2018 via telephone, and a follow-up e-mail was also sent containing a technical bulletin #TB000079, entitled Product Correction Notice, dated March 14, 2018. The notice identified the affected devices and the reason for the recall. Customers were asked to please discontinue use of and dispose of all remaining kits from the affected lots, and document the disposal by completing the 'Certificate of Kit Disposal.' Replacement kits will be provided.

Root Cause

Under Investigation by firm

Distribution Pattern

Distributed to IL, TX, CA, SC, and DE.