LeverEdge Contrast Injector: 35 cc syringe with ISO 594 female luer lock - Product Usage: intended to be used to perfuse contrast medias into the vessels for angiographic procedures, Catalog #: 4100-00.

FDA Recall Z-1529-2021 · Posted None

Recall StatusTerminated
Product CodeDXT
Date Initiated2021-04-06
Date Terminated2022-12-29
Product Quantity616 units

Reason for Recall

The firm received a complaint regarding particulate found in the packaging foam that could contaminate the sterile field.

Firm Action

On 04/01/2021, LeMaitre issued an Urgent Medical Device Removal notice via letter informing customers that particulate contamination has been detected from the packaging foam with these devices.

Root Cause

Process control

Distribution Pattern

Worldwide distribution - US Nationwide distribution in the states of AL, CA, FL, IL, IN, KY, MA, MO, NY, OH, TN, TX and the countries of Bahamas, Belgium, France, Germany, Hong Kong, Spain.