Terminated

Acorn 180 Curved Stairlift

FDA Recall Z-1530-2016 · Posted None

Recalling FirmAcorn Stairlifts Inc
Recall StatusTerminated
Product CodePCD
Date Initiated2016-02-17
Date Terminated2019-01-29
Product Quantity1,290 USA and 9,607 Foreign

Reason for Recall

Aluminum rivets holding the base squab to the framer under prolonged customer use were seen to elongate and subsequently break.

Firm Action

Customers were notified by phone calls, documents mailed with a Rationalization Action plan to follow. Contact information 1-888-563-0410 x1500 or 1400

Root Cause

Component change control

Distribution Pattern

Worldwide Distribution - USA (nationwide) Distribution to the states of : AL, AZ, CA, CO, DE, FL, GA, IL, IN, KS, KY, LA, MA, ME, MD, MI, MO, NH, NJ, NY, NC, OH, PA, RI, SC, TN, TX, UT, VT, VI, WA, WV, WI, and District of Columbia and to the countries of UK and Germany.