Medtronic Abre Venous Self-expanding Sent System, Catalog number REF AB9U16080090 - Product Usage: intended for use in the iliofemoral veins for the treatment of symptomatic venous outflow obstruction.
FDA Recall Z-1532-2021 · Posted None
Advertisement
Recalling FirmMedtronic Inc.
Recall StatusTerminated
Product CodeQAN
Date Initiated2021-03-16
Date Terminated2024-04-15
Product Quantity23 devices
Reason for Recall
Incorrect size printed on the device; packaging is labeled correctly.
Firm Action
Beginning 16-Mar-2021, Medtronic initiated verbal communications to consignees in possession of affected units. Consignees were advised to quarantine all affected units within their possession, pending retrieval.
A written Medical Device Correction letter was mailed on 03/30/2021 via UPS 2-day delivery to consignees who have received affected product according to Medtronic records. Consignees will be asked to complete and return a confirmation certificate, confirming they have received the notification. Consignees are also asked to return all unused, affected product within their
possession.
Root Cause
Process control
Distribution Pattern
US Nationwide distribution in the states of NC, CO, CA, AZ, MT, MO.