POLYSIN (Polyglycolic Acid) suture, HSM9 5-0 BR PGA Part Number G503N

FDA Recall Z-1534-2024 · Posted 2024-04-12

Recalling FirmSURGICAL SPECIALTIES
Recall StatusOpen, Classified
Product CodeGAM
Date Initiated2024-03-26
Product Quantity2,459 units

Reason for Recall

Complaints have been received with allegations of Incorrect Needle Curvature.

Firm Action

On March 26, 2024 customers were sent URGENT MEDICAL DEVICE RECALL letters via email. The letters instructed customers to immediately examine your inventory and quarantine product subject to recall. In addition, if you may have further distributed this product, please identify your customers and notify them at once of the product recall. Your notification to your customers may be enhanced by including a copy of this recall notification letter and the acknowledgement form for product return. We kindly request the return of any affected product that may still be in your stock inventory, and we will replace the product at no charge. If you have any questions, contact [email protected].

Root Cause

Process control

Distribution Pattern

Domestic: AZ, CA, IL, MA, MD, MN, NJ, PA and TX. International: New Zealand.