Dimension Vista CHEM 3 calibrator is an in vitro diagnostic product for the calibration of ammonia (AMON), Cardon Dioxide (CO2) and Ethyl Alcohol (ETOH) on the Dimension Vista System
FDA Recall Z-1535-2010 · Posted 2010-05-04
Advertisement
Recalling FirmSiemens Healthcare Diagnostics, Inc.
Recall StatusTerminated
Product CodeJIX
Date Initiated2009-10-23
Date Terminated2010-06-22
Product Quantity1753 cartons
Reason for Recall
patient results are depressed
Firm Action
An "Urgent Field Safety Notice" dated October 2009 was issued by Siemens to all customers. The notice described the product, problem and actions to be taken by the customers. The customers were instructed to recalibrate the AMON method (AMON Flex reagent cartridge, K3019) with Vista Chem 3 Claibrator lot 9ED058 using the revised bottle values provided in the notice and to please consult with their laboratory director regarding re-evaluation of previous results. The customers should forward this notification to anyone to whom they may have distributed this product.
If you have any technical questions regarding this information, please contact Siemens Technical Solutions Center at 1-800-441-9250.
Root Cause
Other
Distribution Pattern
Worldwide distribution: USA and Australia, Canada, Japan, New Zealand, South Korea, and Switzerland.