iChemVELOCITY Strips, Catalog 800-7212
FDA Recall Z-1535-2020 · Posted None
Advertisement
Recalling FirmBeckman Coulter, Inc.
Recall StatusTerminated
Product CodeKQO
Date Initiated2020-02-17
Date Terminated2021-05-13
Product Quantity10822 vials
Reason for Recall
The Leukocyte and Compensation pads were switched on some vials of lot number 7212269M of the iChem VELOCITY Strips.
Firm Action
An Urgent Medical Device Recall letter will be sent during the week of 02/17/2020 to the affected customers via email and mail. Customers are advised to discard the affected product and order a replacement from BEC.
Root Cause
Under Investigation by firm
Distribution Pattern
Distribution in US including Puerto Rico, Canada, and Taiwan
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.