Terminated

Anti EpCam PE - Analyte Specific Reagent. Anti-EpCAM is directed against human epithelial adhesion molecule that has been identified as a 40 -kilodalton (kd) protein in Western Blotting.

FDA Recall Z-1536-2014 · Posted 2014-05-01

Recall StatusTerminated
Product CodeMVU
Date Initiated2014-04-04
Date Terminated2015-09-28
Product Quantity12 for lot 3295974; 10 of lot 3259822

Reason for Recall

Two lots of Anti EpCAM PE-Analyte Specific Reagent are mislabeled - the actual antibody present in the vial is CD9 PE.

Firm Action

Customer product recall notification letters were sent to all consignees on April 4, 2014 by certified mail.

Root Cause

Under Investigation by firm

Distribution Pattern

Worldwide Distribution-USA including the states of CA, TX, FL, MO, NY, MI, MO, and AZ and Seoul, Korea.

Official Source

View FDA Recall →