Terminated

WET-FIELD OSHER THERMODOT MARKER, 25GA FINE TIP, CURVED; Part number 221270. Designed to create an ink-free, fine dot on the eye that can be used as a precise reference point for a variety of procedures.

FDA Recall Z-1539-2018 · Posted None

Recall StatusTerminated
Product CodeGEI
Date Initiated2018-02-26
Date Terminated2019-01-28
Product Quantity110 total products

Reason for Recall

Incorrect tip was used during manufacturing of the Osher Thermodot Marker. The markers have a needle tip instead of the proper flat tip.

Firm Action

On February 26, 2018, Beaver Visitec International Inc. emailed Product Recall Notices to affected customers. The letter identified the affected device and the reason for the recall. Customers were instructed to destroy affected product, provide evidence of the destruction, and complete a customer response form. Customers will receive a credit. Customers with questions regarding the recall may contact Customer Service at (866)906-8080.

Root Cause

Process control

Distribution Pattern

Distributed to the states of OH, NY, FL, MN, and KS.