Terminated

VersaTREK Myco AS.

FDA Recall Z-1540-05 · Posted 2005-09-14

Recall StatusTerminated
Product CodeMDB
Date Initiated2005-08-08
Date Terminated2007-07-21
Product Quantity29 boxes containing 5 vials in each box.

Reason for Recall

Invitro diagnostic test kit used for testing to confirm the presence of tuberculosis contains interfering bioburden which can cause false positive results.

Firm Action

The recalling firm phoned all 13 consignees that the product was distributed to, and sent a recall letter to all 13 consignees, dated August 8, 2005. All 13 consignees have responded back to the recalling firm and have discarded any remaining product that was on hand.

Root Cause

Other

Distribution Pattern

The product was distributed to consignees located in AL, IA, IL, MD, NJ, OH, PA, Washington D.C. and to two foreign distributors: I.L. Diagnostics, Lago Victoria No. 80, Col. Granada, Mexico D.F., 11520; and Trivitron Diagnostics, #15 4th Street, Abhiramapuram, Chennai, India 600-018.