Recall StatusOpen, Classified
Date Initiated2025-03-04
Product Quantity1493 devices
Reason for Recall
There is a software issue that can permanently disable communication with an implantable neurostimulator (INS) when a specific intraoperative programming sequence is used.
Firm Action
An "URGENT: MEDICAL DEVICE CORRECTION" notification letter dated March 2025 was sent to customers. The letter described the product, problem and actions to be taken.
Required Actions:
When utilizing any of the lead options mentioned above, avoid using the Switch Device feature to transfer data from the Model A71300 Stimulation Trialing CP App to the Model A71400 Stimulation RC CP App. As an alternative, manually enter the patient and lead information using the Model A71400 Stimulation RC CP App, ensuring not to utilize the Switch Device feature.
Inform relevant medical staff about the potential of this issue and the steps to mitigate it.
Ensure that backup INS devices are readily available for scheduled surgeries.
During the implant procedure, check the system integrity before securing the neurostimulator in place as described in the Inceptiv/Intellis Pro Implant Manual.
Please complete and return the customer confirmation form enclosed in this letter acknowledging that you have received this information.
Please contact Medtronic Technical Services at 1-800-707-0933 if you have any questions
regarding this communication.
Root Cause
Under Investigation by firm
Distribution Pattern
Worldwide Distribution: US (Nationwide) and OUS (Global) to countries of: Austria, Belgium, Bulgaria, Canary Islands, Cyprus, Croatia, Czech Republic, Denmark, Finland,
France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Japan, Luxembourg, Malta, Netherlands, Northern Ireland, Norway, Poland, Portugal, Puerto Rico, Slovakia, Slovenia, Spain, Sweden, Switzerland, Turkey, and United Kingdom.