Terminated

12-Instrument Sterilization Tray; Product Usage: The 12-Instrument Tray is used to hold and protect surgical instruments during the sterilization process. Specifically, this tray is used for the Conquest Manual Instrument line. The tray consists of an interlocking tray and lid which are perforated to allow for the passage of sterilizing agent from outside the tray to the inside of the tray where the devices are held. The tray features silicon device holders that secure devices during the sterilization process.

FDA Recall Z-1544-2014 · Posted 2014-05-02

Recalling FirmStryker Endoscopy
Recall StatusTerminated
Product CodeGEI
Date Initiated2014-04-16
Date Terminated2014-08-18
Product Quantity2591 trays

Reason for Recall

Gravity sterilization parameters provided in the 12 Instrument Sterilization reprocessing guide did not fully consider all worst-case scenarios during validation testing.

Firm Action

Stryker sent an Urgent Device Corrections letters dated April 18, 2014 to all US customers via certified mail. Stryker Sales representative and Stryker international distributors were notified with the letter and response card by e-mail. International sites are responsible for notifying customers. The letter identified the affected product, problem, potential hazards and actions to be taken. Customers were instructed to complete and sign the enclosed Acknowledgement of Receipt form and scan and e-mail to [email protected]. For questions call 1-800-624-4422 option 3 or via email at [email protected].

Root Cause

Reprocessing Controls

Distribution Pattern

Worldwide distribution