Terminated

Alcohol Control sold by Pointe Scientific; Catalog # 7-A7504-CTL.

FDA Recall Z-1546-06 · Posted 2006-09-30

Recall StatusTerminated
Product CodeJIX
Date Initiated2006-06-09
Date Terminated2007-11-20
Product Quantity44 units

Reason for Recall

Open vial stability-Potential microbial growth in the alcohol standard and possible unexpected QC changes in the ammonia portion of the control due to the lack of validation for long-term open vial stability.

Firm Action

Consignees were notified via recall letter dated 6/9/06 and a second letter dated 6/21/06 extending the recall to additional products/lots, to cease using the products, conduct a recall to the user level, destroy products on hand and to notify the firm via the response form of the amount destroyed for replacement purposes.

Root Cause

Other

Distribution Pattern

Worldwide distribution ---- including USA and countries of Canada, Ecuador, Germany, Greece, India, Mexico, Peru, Poland, Portugal, Trinidad, United Arab Emirates and United Kingdom.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.