Terminated

Medi-Vac(TM) Flex Advantage(TM) Suction Canister Liner, Material Numbers: a) REF 6561-910C b) REF 6561-916K c) REF 6561-920C To remove blood and/or body fluids. These fluids may be suctioned during surgical procedures or other applications requiring the removal of built up fluids. The suction canisters are used in operating rooms but are also used outside the surgical field as well as post-operatively within patient suites. The collection containers hold the fluids until they are ready for proper disposal. Sources of vacuum range from wall suction (central vacuum piped throughout the healthcare facility) to liposuction devices (portable suction pumps).

FDA Recall Z-1546-2018 · Posted None

Recall StatusTerminated
Product CodeGCX
Date Initiated2018-03-28
Date Terminated2018-11-29
Product Quantity3600 units

Reason for Recall

products may be missing a one-way valve on the fluid patient port.

Firm Action

Cardinal Health sent an Urgent Product Recall letter dated March 28, 2018, Please take the following actions: 1. INSPECT your inventory for the affected product code and lot number (listed on package) 2. SEGREGATE and QUARANTINE all on-hand product 3. RETURN the enclosed acknowledgment form via fax to 847-689-9101 or 614-652- 9648, whether or not you have affected product, indicating the product code, lot and quantity of product you've quarantined 4. CONTACT Customer Service at 800-964-5227 to arrange for return and credit of any affected product . For further questions, please call (800) 292- 9332.

Root Cause

Process control

Distribution Pattern

US Distribution to the state of MA

Official Source

View FDA Recall →