Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

Medline Sterile 3mL Syringes with 21GA x 1 Inch Needle 80 each REF DYND3ML21G1A lot 18JBH655 (01) 1088438982153.

FDA Recall Z-1546-2020 · Posted None

Recall StatusTerminated
Product CodeLRO
Date Initiated2020-02-10
Date Terminated2023-02-13
Product Quantity3,690 units

Reason for Recall

Voids were discovered in the packaging seal and loss of sterility on the outside of the contents within the Tyvek Form-Fill-Seal packaging.

Firm Action

Medline notified customers on February 18, 2020 via "Urgent Recall Immediate Action required" letter. The consignees were instructed to complete the response form and return the product. The recall letter identified the affected product and lot numbers. Reason for the correction was also noted in the recall letter.

Root Cause

Packaging

Distribution Pattern

US Nationwide: MA, LA, IL, MO, MI, MN, NJ, KS, GA, NY, PA, SD, CT, TN, MD, OH, VA, FL, TX, KY, CO, IN, AL, CA, MT, OR, AK, AL, DC, IA, ME, NV, NE, SC, WV, WY, RI, NC, AZ, UT, DE and WI.

Official Source

View FDA Recall →