TC-PLUS(TM), Primary Fixed Tibial Component VKS Right, SIZE 4 RIGHT, STERILE R, REF 12 000 024, S&N 7500279, Smith & Nephew Orthopaedics AG Product Usage: Tibial knee prosthesis

FDA Recall Z-1547-2016 · Posted 2016-04-21

Recalling FirmSmith & Nephew, Inc.
Recall StatusTerminated
Product CodeJWH
Date Initiated2016-03-11
Date Terminated2017-04-10
Product Quantity33 units

Reason for Recall

The Affected Devices may contain a dimensional deviation that could potentially result in the surgeon being unable to seat the polyethylene (PE) Insert on the affected VKS tibial base plate during surgery.

Firm Action

The firm, Smith & Nephew, issued an "URGENT FIELD SAFETY NOTICE: Medical Device Field Safety Corrective Action/Recall" letter dated 3/10/2016 to its local QA/RA personnel/customers in the affected countries by email on 03/11/2016. The letter described the product, problem and actions to be taken. The customers were instructed to locate and quarantine affected unused devices immediately; return quarantined product to your national Smith & Nephew agency/distributor; complete and return the Return Slip via fax to your national Smith & Nephew agency/distributor; and pass on this safety information to all those who need to be aware in your organization. The local QA/RA staff will then notify end customers according to local procedures. In Germany as well as in Switzerland, sales representatives will inform the affected customers in-person and retrieve affected devices from their stock. If you have any questions, call 978-749-1330 or email: [email protected]

Root Cause

Process control

Distribution Pattern

International Distribution only in countries of: Germany and Switzerland.