2008 Series Hemodialysis Machines: 2008Kathome - 190895 2008T GEN 2 BIBAG WITHOUT CDX; 190908 2008T HD SYSTEM WITH CDX, CANADA; 190713 2008T HEMODIALYSIS SYS., with CDX; 190766 2008T HEMODIALYSIS SYSTEM W/BIBAG; 190858 2008T HEMODIALYSIS SYSTEM WITHOUT CDX The Fresenius 2008K is indicated for acute and chronic dialysis therapy.

FDA Recall Z-1549-2015 · Posted 2015-04-29

Recall StatusTerminated
Product CodeKDI
Date Initiated2015-02-20
Date Terminated2021-05-18
Product Quantity133,751 in total (US: 130,337 and OUS: Canada 1,201 and Mexico 2,213)

Reason for Recall

The Acetate value for GranuFlo on the Select Concentrate screen should be 8mEq/l but it is displayed as 4mEq/l. The correct value is shown on the dialysate screen.

Firm Action

The firm, Fresenius Medical Care, sent an "URGENT - MEDICAL DEVICE CORRECTION" letter dated February 19, 2015 to its Consignees via certified mail on February 26, 2015. The letter describes the product, problem and actions to be taken. The consignees were instructed to sign and return the attached "Reply Form" via fax to: 781-699-9769 Attn: Quality Department or Scan and E-mail to: [email protected] to confirm that they received and understand the customer notification. The firm will be providing a software update and stated that in the near future, your local clinic will be contacting you about installing the software update. If you have any questions pertaining to this Customer Notification, please contact your local clinic, 800-227-2572 or the Corrections and Removals Manager at 800-662-1237.

Root Cause

Software design

Distribution Pattern

Worldwide Distribution: US (Nationwide) and countries of: Canada and Mexico.