Peloris Rapid Tissue Processor; Leica Biosystems Melbourne Pty Ltd, Mount Waverly, 3149 VIC, Australia; catalog number 26.0001 and 26.0005
FDA Recall Z-1551-2010 · Posted 2010-05-07
Advertisement
Recalling FirmLeica Microsystems, Inc.
Recall StatusTerminated
Product CodeIEO
Date Initiated2010-02-09
Date Terminated2011-06-30
Product Quantity574 units
Reason for Recall
Confusion during manually changing reagents, general users running unvalidated protocols, and wax valve leakage have contributed to tissue damage during processing, leading to rebiopsies.
Firm Action
Leica Microsystems sent Medical Device Correction letters dated 2/9/10 to the affected customers via first class mail on the same date, informing them of the potential for tissue damage during processing and outlining the software upgrade being made to minimize the risk of tissue damage. Customers were informed that their local Leica Microsystems representative will contact them to arrange the software upgrade and to provide training on the use of the new software. The letters also provided the necessary steps required to avoid tissue damage during processing while they are awaiting the software upgrade. The accounts were asked to complete the attached Acknowledgement Form, listing the serial numbers of the units at their facility and fax it back to Leica at 1-847-607-3147.
Additional questions should be addressed to the firm's office of Regulatory Affairs and Quality Assurance.
Root Cause
Other
Distribution Pattern
Worldwide distribution -- U.S. (nationwide), Canada, Australia, Europe and Asia.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.