Terminated

Maquet Getinge-BEQ-TOP 40700 HUNTINGTON Material:701062899

FDA Recall Z-1551-2019 · Posted 2019-05-31

Recall StatusTerminated
Product CodeDWF
Date Initiated2019-04-18
Date Terminated2025-07-28

Reason for Recall

Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path between the closed port and the non-vented cap therefore the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured

Firm Action

Maquet Getinge issued recall notification letter on April 18, 2019 via FedEx 2-day delivery identifies issue, health risk and action to take: quarantine and return to Maquet/Getinge recalled product. Complete response form and return to acknowledge this recall. Questions contact Getinge Customer Service at 1 888 627 8383 (press option 2, then option 2) between the hours of 8 AM and 6 PM Eastern Standard Time. On 3/26/20 firm expanded recall to include an additional lot for one consignee,product: "BEQ TOP 25003 3/8 3/8 up to 7 lpm", item number 701063255R02, lot 3000087051.

Root Cause

Device Design

Distribution Pattern

Nationwide