Terminated

ADVIA Centaur XP Immunoassay system, an automated in vitro diagnostic analyzer. Catalog # 078-A011-03 - ADVIA Centaur XP, Catalog # 078-A011R03 - ADVIA Centaur XP refurb.

FDA Recall Z-1553-2014 · Posted 2014-05-13

Recall StatusTerminated
Product CodeJJE
Date Initiated2014-02-18
Date Terminated2017-03-06
Product Quantity5,900 units total (1,679 domestically & 4,221 internationally)

Reason for Recall

Instrument Cover Gas Spring failures. Over time, the gas spring may lose its effectiveness and fail to support the cover in its full or partially open position. This may lead to the cover falling.

Firm Action

On February 18, 2014 Siemens Healthcare Diagnostics distributed Urgent Medical Device Correction notices to their customers via courier service (Urgent Medical Device Correction/). Upon receipt of the communication, customers are asked to follow "Actions to be Taken by the Customer". If the cover is unable to stay partially raised, immediately contact your Siemens technical support representative to arrange for a service. Complete and return the Field Correction Effectiveness Check attached to this letter within thirty (30) days by faxing to (302) 631-7597. Customers with questions should contact their Siemens technical support representative.

Root Cause

Device Design

Distribution Pattern

Nationwide and internationally