Zimmer Dermacarriers Skin Graft Carriers, S.G. Carrier 16IN length, Rx, latex free, sterile, Zimmer, Dover, Ohio; REF 00-7716-000-10. Single use device for use during skin grafting procedures.

FDA Recall Z-1555-2008 · Posted 2008-07-30

Recalling FirmZimmer Inc.
Recall StatusTerminated
Product CodeFZW
Date Initiated2008-04-03
Date Terminated2009-12-17
Product Quantity99345 of all units

Reason for Recall

Sterility compromised. Lack of assurance of sterility, as the packages may not have been sealed on the manufacturer's end.

Firm Action

Consignees were notified via letter (Urgent: Medical Device Correction) dated 4/3/08 to identify their stocks on hand, verify package sealing, to discard any non-sealed packages, mail the Business Reply Card regardless if there are any remaining affected products.

Root Cause

Packaging process control

Distribution Pattern

Worldwide Distribution --- including USA and countries of Algeria, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Czech Republic, Denmark, Dubai, Ecuador, England, Finland, France, Germany, Ghana, Greece, Guatemala, Hong Kong, Hungary, Iceland, India, Iraq, Israel, Italy, Japan, Jordan, Kenya, Largos, Lebanon, Lithuania, Malta, Mexico, Morocco, Netherlands, Nigeria, Panama, Peru, Poland, Saudi Arabia, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Tanzania, Thailand, Tunis, Turkey and Venezuela.