Venue Go Models: R2 (Model Number H45181VG), R3 (Model Number H45193VG), R4 (Model Number H45114VG), R5 (Model Numbers H45135VG, H45135VGW), R3 Surgery for China (Model Number H45113VGSU), R3 PeriOP for China (Model Number H45113VGPO), R3 Expert for China (Model Number H45113VGEX ), R3 Pro for China (Model Number H45113VGPR), R3 Performance for China, R3 Intervention for China, R3 Surgery for China (Model Number H45113VGSU) diagnostic ultrasound systems
FDA Recall Z-1555-2025 · Posted 2025-04-11
Advertisement
Recalling FirmGE Medical Systems, LLC
Recall StatusOpen, Classified
Product CodeIYN
Date Initiated2025-03-06
Product Quantity12,808 total devices
Reason for Recall
The battery for certain ultrasound systems can potentially develop an internal failure which could result in smoke or fire.
Firm Action
An URGENT MEDICAL DEVICE CORRECTION notification letter dated 3/6/25 was sent to customers.
Actions to be taken by Customer/User:
You can continue to use your Venue Go and Venue Fit system after following the actions below:
1) Turn the system Off and unplug the system from the wall socket.
2) Remove the system batteries by following instructions provided in the User Manual Chapter 3. These instructions are also provided in this letter in Appendix A for Venue Go and in Appendix B for Venue Fit.
3) Check the date of manufacture of the system batteries by looking at the label on the bottom of each battery (See Figure 1). The date of manufacture is displayed as YYYY-MM for year and month of manufacture.
NOTE: Venue Go has two batteries and Venue Fit has one battery
4) Calculate the age of each of the system batteries.
5) If any of the batteries are more than 2 years old:
a. With the system unplugged, re-insert batteries, turn the system On and leave the system On to discharge the batteries.
b. Once the battery capacity indicator shows two segments or less (See Figure 2), turn the system Off and remove the batteries per step 2.
c. Dispose of batteries according to local regulations.
NOTE: After battery removal, you can continue to use the system with use of AC power cord. The system is fully functional without the battery.
6) If the batteries are less than 2 years old, you may continue to use the battery until the battery reaches 2 years old, at which point, repeat instructions in step 5.
Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions.
Please retain this document for your records.
Please complete and return the attached acknowledgement form to [email protected].
GE HealthCare will correct all affected products at no cost to you.
A GE HealthCare representative will contact you to arrange for the correction.
After the correction has been implemented, please be sure to destroy
Root Cause
Under Investigation by firm
Distribution Pattern
Worldwide - US Nationwide distribution.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
K240053Venue GoGE Medical Systems Ultrasound and Primary Care Diagnostics · 2024-06-10K220800Venue GoGE Medical Systems Ultrasound and Primary Care Diagnostics · 2022-06-21K202233Venue GoGE Medical Systems Ultrasound and Primary Care Diagnostics · 2020-11-18K183362Venue GOGE Medical Systems Ultrasound and Primary Care Diagnostics · 2019-01-28