Eyesuite 9.3.1 software, LENSTAR, MODEL LS900: ac-powered slit-lamp biomicroscope.

FDA Recall Z-1560-2021 · Posted None

Recalling FirmHaag-Streit USA Inc
Recall StatusOpen, Classified
Product CodeHJO
Date Initiated2021-04-15

Reason for Recall

Examination data and patient name may be mixed up when printing or generating a pdf with the interface at Pacific Coast and Laser Institute (PCLI).

Firm Action

On 04/06/2021, Haag-Streit issued an Urgent Medical Device Correction notice to their customer via letter notifying them Eyesuite 9.3.1 software may mix up examination data and patient name when printing or generating a PDF.

Root Cause

Software design

Distribution Pattern

US Nationwide distribution in the state of WA.