PM1000 Flow Selector - An accessory to an emergency ventilator (resuscitator). The selector is attached to a flowmeter and is used to direct the outlet flow from any standard flowmeter to any one of the three outlets it provides.

FDA Recall Z-1561-2010 · Posted 2010-05-09

Recall StatusTerminated
Product CodeBTM
Date Initiated2010-03-05
Date Terminated2010-12-13
Product Quantity19,518

Reason for Recall

The on/off outlet is mislabeled so that when a user/operator turns selector to "on", it actually may be turned "off"

Firm Action

The recalling firm issued Customer Alert Product Recall letters on 3/5/10 to inform users of the problem and how to identify affected product.

Root Cause

Labeling mix-ups

Distribution Pattern

The products were shipped to distributor and medical facilities nationwide. The products were also shipped to Taiwan, Germany, Hong Kong, Canada