DigitalDiagnost C50 1.1
FDA Recall Z-1563-2022 · Posted 2022-09-01
Advertisement
Recalling FirmPhilips North America Llc
Recall StatusOpen, Classified
Product CodeKPR
Date Initiated2022-02-25
Product Quantity11 systems (updated 1/3/2023)
Reason for Recall
Following a system restart where the Table Height 2 (TH2) is the default setting and then switched to Wallstand VS2, operators will see an incorrect orientation of the image on the first examination due to an issue in the firmware of the Wallstand VS2 board. The system will rotate the amplimat field selection by 90 degrees. The wrong amplimat field selection may cause an incorrect dose of radiation to occur. Additionally, the anatomic position markers may become mispositioned and could potentially be associated with the opposite side of the anatomy.
(Added 1/3/2023), Expansion of the root cause to include incorrect amplimat field selection by 90 degrees following the first exposure using Wallstand VS2 and a subsequent system restart. Previously, this issue only occurred following a system restart where the Table Height 2 (TH2) is the default setting.
Firm Action
Philips notified customers of the issue via an Electronic Product Radiation Defect notification letter dated July 12, 2022. Philips will schedule an appointment with customers to install the software update at no cost to customers. For questions contact Philips Customer Care Solutions Center 1-800-722-9377.
Root Cause
Software Design Change
Distribution Pattern
Nationwide Distribution including Puerto Rico
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.