Magellan Robotic Catheter Accessory Kits, Model number MACC 50 and MACC80. Cardiology: The Hansen Medical Magellan System is a robotic control system that enables positioning and visualization of a Magellan Catheter and an off-the-shelf guide wire while allowing a physician to remain seated and away from the x-ray radiation source. The Accessory Kits contain disposable components that keep the Magellan Catheters and 3rd-party guide wires performing smoothly during system operation.

FDA Recall Z-1564-2016 · Posted None

Recalling FirmHansen Medical Inc
Recall StatusTerminated
Product CodeDXX
Date Initiated2016-03-25
Date Terminated2016-05-06
Product Quantity8 kits

Reason for Recall

Firm's quarterly dose audit testing showed that one batch was radiated below the augmented dose and therefore did not meet the augmented dose of 28.4 kGy.

Firm Action

Hansen Medical sent an Urgent Notice Product Recall notification dated March 28, 2016, to all affected customers. The notification identified the product, the problem, and the action to be taken by the customer. Customers were instructed to provide the product to the Hansen Medical field representative to assure credit appropriately applied to their account. Customers with questions were instructed to contact their local Hansen Medical representative or call 650-404-5581.

Root Cause

Under Investigation by firm

Distribution Pattern

Nationwide Distribution to OH, TX. FL, CA and GA.