Automated peritoneal dialysis (APD) cycler
FDA Recall Z-1565-2016 · Posted None
Advertisement
Recalling FirmBaxter Healthcare Corp.
Recall StatusTerminated
Product CodeFKX
Date Initiated2016-02-16
Date Terminated2016-12-23
Product Quantity10 units
Reason for Recall
Baxter Healthcare Corporation is sending this communication to inform you of incomplete instructions in the AMIA Automated Peritoneal Dialysis (PO) System Clinician Guide for calculating the total recommended solution therapy volume. Specifically, the instructions do not specify the need for an extra 200 mL of PD solution in order to prime the patient line and for air purge operations.
The tot
Firm Action
An Urgent Device Correction communication was sent to the one affected customer via U.S.P.S., first class mail on 02/11/2016.
A Clinician Guide Addendum was created to instruct clinicians of the extra 200 mL of PD solution required to prime the patient line and for air purge operations. This Addendum was provided as an attachment to the Urgent Device Correction communication letter.
Root Cause
Under Investigation by firm
Distribution Pattern
US Consignee: NY ***Foreign Consignee: Ecuador
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.