Terminated

AxSYM Ultrasensitive hTSH II Master Calibrators Microparticle Enzyme Immunoassay that determines amount of human thyroid stimulating hormone in human serum and plasma.

FDA Recall Z-1566-2008 · Posted 2008-07-24

Recall StatusTerminated
Product CodeJIS
Date Initiated2008-03-06
Date Terminated2012-01-25
Product Quantity882 total units

Reason for Recall

High Control value out of range after calibration. --AN INCREASED FREQUENCY OF HIGH CONTROL VALUES BEYOND THE UPPER LIMIT OF THE RANGE SPECIFIED BY THE PACKAGE INSERT.

Firm Action

The firm issued Product Recall letters dated February 22, 2008, requesting remaining units of the affected product be discontinue using and destroyed. A Customer Reply Immediate Action Required letter was enclosed to indicate acknowledging receipt of the product recall letter and to provide information on the number of kits destroyed before March 7, 2008 and faxed to 1-800-777-0051.

Root Cause

Nonconforming Material/Component

Distribution Pattern

Worldwide Distribution --- USA including states of AZ, CA, CO, FL, GA, ID, IL, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NY, OH, OK, PA, RI, SC, SD, TN, TX, WA, WI, WV, and WY, and Puerto Rico, and countries of Canada, Mexico, Brazil, Chile, Germany, Japan, and Hong Kong.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.