Terminated

TRUE Dilatation Balloon Valvuloplasty Catheter, REF Numbers: 02045-11, 02245-12, 02445-12, and 02645-13; Sizes: 20mm x 4.5cm, 22mm x 4.5cm, 24mm x 4.5cm, and 26mm x 4.5cm; Intended to be used for balloon aortic valvuloplasty.

FDA Recall Z-1566-2013 · Posted 2013-06-21

Recalling FirmLoma Vista Medical
Recall StatusTerminated
Product CodeOZT
Date Initiated2013-03-18
Date Terminated2013-07-17
Product Quantity2379 units

Reason for Recall

Slow deflation time: Slower than normal balloon deflation time of greater than 10 seconds. The balloon normally deflates in less than 5 seconds. The device performance specification for deflation time is 10 seconds maximum.

Firm Action

The firm sent an "URGENT VOLUNTARY DEVICE REMOVAL" letter dated March 18, 2013 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers.

Root Cause

Process design

Distribution Pattern

Worldwide Distribution-US Distribution in New Jersey and the countries of Germany, Switzerland, Italy, Norway, The Netherlands, Spain, Portugal, Ireland, and Austria.