Centricity Universal Viewer Software Versions 7.0 through 7.0 Sp2.0.1, a device that displays medical images (including mammograms) and data from various imaging sources, Model Numbers 5826659-027, 5826659-029, 5826659-032, 5826659-033, 5826659-028, 5826659-034, 5826659-030, 5826659-019, 5826659-022, 5826659-026, 5826659-023, 5826659-031, 5826659-025, 5826659-002, 5826659-021, 5826659-037, 5826659-014, 5826659-016, 5826659-013
FDA Recall Z-1569-2026 · Posted 2026-03-16
Recalling FirmGE Medical Systems, LLC
Recall StatusOpen, Classified
Product CodeLLZ
Date Initiated2026-01-30
Product Quantity1568 units
Reason for Recall
There is a potential cybersecurity vulnerability affecting certain versions of Centricity Universal Viewer. User login credentials may be exposed on the local client workstation, which could allow an unauthorized individual to potentially impact system availability and/or manipulate data.
Firm Action
An URGENT MEDICAL DEVICE CORRECTION notification letter dated 1/30/26 was sent to customers.
Actions to be taken by Customer/User
Pending corrections from GE HealthCare, you can continue using your device by following the instructions below:
Ensure workstations have appropriate security controls per product manuals.
Implement network account authentication by using Active Directory/LDAP services for user management.
If network account authentication is not possible at your site, please contact GE HealthCare Service to request temporary steps to mitigate the issue.
Ensure all potential users in your facility are informed of this safety notification.
Please retain this document for your records.
Please complete and return the attached acknowledgement form to [email protected].
GE HealthCare will correct all affected products at no cost to you. A GE HealthCare representative will contact you to arrange the correction.
If you have any questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.
Root Cause
Software design
Distribution Pattern
Worldwide - US Nationwide distribution.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.