ACUSON X700 Ultrasound System with software version 1.0.04. Model numbers: 10658844 ACUSON X700 Ultrasound System 10658846 - ACUSON X700 Ultrasound System (Russia) 10658845 - ACUSON X700 Ultrasound System (Korea) Radiology: The Siemens ACUSON X700 ultrasound imaging system is intended for the following applications: Cardiac (Adult, Pediatric), Transesphageal (Cardiac), Intracardiac, Cerebrovascular, Peripheral Vessel, Abdominal, Renal, Fetal, Abdominal, Intra-operative, Pediatric, Small Organ, Neonatal Cepahalic, Adult Cephalic, Orthopedics, Musculo-skeletal Conventional, Musculo-skeletal Superficial, Pelvic, Obstetrical, Gynecological and Urological applications. The system also provides for the measurement of anatomical structures and for analysis packages that provide information that is used for clinical diagnosis purposes.
FDA Recall Z-1572-2016 · Posted None
Advertisement
Recalling FirmSiemens Medical Solutions USA, Inc.
Recall StatusTerminated
Product CodeIYN
Date Initiated2015-09-05
Date Terminated2016-04-27
Product Quantity4
Reason for Recall
Due to a communication error between the software and V5Ms transducers rotation function, ACUSON X700 ultrasound systems at software versions 1.1.04, display an IMG_15 error message causes the system to lock up, which requires a reboot to recover system operation.
Firm Action
September 5, 2015 we initiated a correction via update program US020/15/S where we visited all four units and updated their software to version 1.0.05
Root Cause
Software design
Distribution Pattern
Distributed in Germany, U.A.E., Hungary and Brazil.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.
Advertisement