SteriSpine PS kit of 2 Multi-Axial Fenestrated Screws, model numbers/REF numbers KITM640, KITM645, and KITM740.
FDA Recall Z-1572-2022 · Posted None
Advertisement
Recalling FirmSAFE ORTHOPAEDICS LLC
Recall StatusCompleted
Product CodeNKB
Date Initiated2022-06-03
Product Quantity8 devices
Reason for Recall
The type of screw printed on the traceability record for the patient label is incorrect, but the type of screw printed on the other labeling is correct.
Firm Action
The recall letter was issued on 6/3/2022 via email to the firm's one U.S. customer. The email included a dealer recall letter, dealer response form, and a template letter for notification of any hospitals who received the product from the dealer. The letter requested the products be quarantined and the dealer should contact the hospitals who received the product to notify them of the incorrect patient labels.
Root Cause
Labeling Change Control
Distribution Pattern
Distribution was made to MD. There was no government/military distribution.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.