Recall StatusOpen, Classified
Date Initiated2023-02-28
Product Quantity932 Reagent Packs
Reason for Recall
APLS IgM reagent kits were packaged with the incorrect conjugate, which could lead to an increase in false-positive and false-negative results.
Firm Action
On February 28, 2023, Bio-Rad Laboratories issued an "Urgent: Medical Device Recall" letter to affected consignees. In addition, to informing consignees about the recall, they ask consignees to take the following actions:
We are requesting that customers perform quality control using the BioPlex APLS IgM
Control Set (Catalog No. 663-2030, UDI GTIN 00847865000642) with each reagent pack ( single pack validation ) of BioPlex 2200 APLS IgM Reagent Pack, Lot No. 301538, until replacement product is provided.
If QC fails (controls below or above the acceptable range), please discontinue use of the reagent pack and dispose of it according to local waste management procedures. If the control values are within the acceptable range, the reagent pack may be used and patient results may be released. If you have distributed or transferred this product to any other Bio-Rad customers, or to any other laboratories from your site, please notify those customers of this field action.
Please complete the attached Customer Response Form and return it to your local Bio-Rad
Technical Support so they can assist you with obtaining replacement reagent packs as needed.
Root Cause
Under Investigation by firm
Distribution Pattern
US Nationwide distribution including in the states of AL, AR, AZ, CA, CT, FL, HI, IL, IN, KS, KY, LA, MI, MO, NC, ND, NE, NJ, NM, NY, OH, OK, PR,SC, TN, TX, UT, VT.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.