Terminated

NeuViz 16 Multi-Slice CT Scanner System PN: 989605858501 a whole body computed tomography X-ray system featuring a continuously rotating X-ray tube and detector array with multi-slice capability up to 16 slices simultaneously.

FDA Recall Z-1573-2016 · Posted 2016-04-27

Recall StatusTerminated
Product CodeJAK
Date Initiated2016-03-24
Date Terminated2016-11-23
Product Quantity37 units

Reason for Recall

The following issues are found in NeuViz 16 systems with software version 1.1.4.21425 and version 1.1.4.21426: 1) During the filming operation on MX 16-slice console software, the clipboard used for copying and pasting images is not cleared between patients. If the operator fails to copy the current patient's images before pasting, a previous patient's image may be present in the clipboard and be

Firm Action

Affected consignees will be notified of the recall via letter on March 24, 2016. Mandatory Field Change Order and FCO kits will be released to all affected systems in June, 2016.

Root Cause

Under Investigation by firm

Distribution Pattern

NC, OH, NE, SC, TX, LA, PR, MO, FL, CT