HANA/PROFx CLASSIC FEMORAL HOOK RIGHT REF 6850-144
FDA Recall Z-1573-2022 · Posted None
Advertisement
Recalling FirmMizuho OSI
Recall StatusOpen, Classified
Product CodeLXH
Date Initiated2021-07-29
Product Quantity13 devices
Reason for Recall
One batch of "regular" femoral hooks were incorrectly etched with the "classic" femoral hook model number.
Firm Action
On or about 07/05/2022, Muzuho OSI Representative started visiting customers to inform them of a labeling error with the Femur Hook, Right 6850-918 which was incorrectly label with part number 6850-144.
Mizuho Representatives were instructed to inform customers:
-Over this summer, their facility should have received a Classic Femur Hook, Right 6850-144, either as part of your Hana Table purchase or as a replacement part. But, in error, may have received the other style of femur hook mislabeled with the part number 6850-144.
-Hand deliver the Customer Notification Letter and Acknowledge Form.
Mizuho Representative are further instructed to:
-Inspect the customers' inventory and swap/replace all incorrectly labeled products with the correctly labeled product.
Root Cause
Process control
Distribution Pattern
U.S Nationwide Distribution .: CA, FL, GA, MI, NE, NJ, TN, UT, and WA O.U.S.: Canada
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.