Terminated

MHI-TM2000 Linear Accelerator System (Software Version 3.5.0 and 3.5.1)

FDA Recall Z-1574-2015 · Posted 2015-05-01

Recall StatusTerminated
Product CodeIYE
Date Initiated2015-04-17
Date Terminated2017-04-11
Product Quantity11 units total

Reason for Recall

The operator console allows users to deliver therapeutic radiation to patients even though a specific communication error relevant to imaging conditions has occurred. In addition, the treatment record (the delivered radiation record) cannot be saved.

Firm Action

MITSUBISHI HEAVY INDUSTRIES, LTD. has issued a Safety Alert for the MHI-TM2000 Linear Accelerator System dated April 10, 2015, due to a software issue. Customers were advised not to deliver therapeutic X-ray if popup warning message show is displayed during treatment. Instructions were provided if therapeutic radiation is delivered during popup warning. Customers with questions should contact the Customer Service Representative by calling 1-800-597-5911, Monday through Friday or email: [email protected]. **CAPA** Correction software for this issue has a tentative availability date of June 2015. Estimated software update completion date would be end of October 2015.

Root Cause

Software design (manufacturing process)

Distribution Pattern

Worldwide Distribution - US including FL, NY, OH, & TX, **Internationally - Belgium, Italy, Korea, France & Germany.