NexGen Augment; 5mm Distal Femoral Augment;Part number 00-5990-036-10. Intended to be used in total knee arthroplasty when there is insufficient bone in the distal femur.
FDA Recall Z-1575-2015 · Posted 2015-05-01
Recalling FirmZimmer, Inc.
Recall StatusTerminated
Product CodeJWH
Date Initiated2015-04-02
Date Terminated2015-09-23
Product Quantity20 units
Reason for Recall
The affected lot is being recalled because it was packaged without the locking screw. The locking screw packaged with the augment is used to fix the augment to the femoral component. This has the potential to delay surgery
Firm Action
On 4/2/2015, URGENT MEDICAL DEVICE RECALL LOT SPECIFIC notifications were sent to the affected distributors with instructions for locating, quarantining, and returning the affected product. All distributors were notified via electronic mail, and distributors with affected inventory were also notified via courier. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. If after reviewing this notification you have further questions or concerns please call the customer call center at 1-877-946-2761 between 8:00 am and 5:00pm EST.
Root Cause
Mixed-up of materials/components
Distribution Pattern
Distributed in the states of IN, MN, FL, and IL, and the countries of Argentina and Germany.