Panorama Patient Monitoring Network Mindray DS USA The Panorama Patient Monitoring can view recal time, store, print, graph and trend patient clinical and demographic data. The Panoram Patient Monitoring Network can sent independent alarm limits for data send by the bedside monitor.
FDA Recall Z-1575-2016 · Posted None
Advertisement
Recalling FirmMindray DS USA, Inc. dba Mindray North America
Recall StatusTerminated
Product CodeMHX
Date Initiated2016-03-09
Date Terminated2018-04-19
Product Quantity1447 units US, 45 units OUS
Reason for Recall
Panorama Central Station including the work station View Station, View Station and the eGateway will revert to the year 2000 when Daylight Savings Time (DST) occurs on March 13, 2016, or under various conditions.
Firm Action
Mindray DS USA Inc. issued a Medical Device Service Notice dated March 9, 2016. The letter identified the product, the problem, and the action to be taken by the customer. Customers requiring technical assistance were instructed to contact Mindray's Technical Support team at 1-800-288-2121. Mindray will address the systems response to DST in a future software update which will be provided to the customer at no cost.
Root Cause
Unknown/Undetermined by firm
Distribution Pattern
Nationwide Distribution
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.